Coil Converter for MagPro Compact
0.00$PRODUCT #PTS3028
Interface unit to be used by MagPro Compact only.
The unit is mounted on the front of MagPro Compact between the stimulator and the coil. The external power control is for coils without controls in the coil handle. Instead, the control is carried out from the external power control.
FEATURES
- Can be used on all coils except
- C-100, C-B60 and MMC-140-II
- *Note: F- and Cool coil types are not recommended for MagPro Compact
- Weight 140g
-
Isolation Transformer MagPro R & X Models, 120V – Magventure
0.00$PRODUCT #MV9016D0031
Required for rTMS system solutions with MagPro stimulators and other devices such as Treatment Chair, Vacuum Pump and Coil Cooler Unit.
Complies with the leakage current requirements of IEC 60601-1-1.
FEATURES
- Weight of transformer 15 kg
- Cable length primary 3 m
- Cable length for MagPRo 2 m
- Height x width x depth 12 x 30 x 23 cm
- Encapsulation Min. 2 mm PC Non flammable; Impact resistant
- Main Voltage Inlets 100V, 120V, 230V
- Outlet for MagPro Fixed cable, 230V
- Auxilliary Outlets 3 pcs. IEC, 230V, total mx. 100VA
-
MEP Monitor 1 Channel EMG Amplifier
0.00$PRODUCT #MV9016C0701
One channel EMG amplifier mounted on the back of the MagPro, optimized for Motor Evoked Potentials.
The MEP Monitor can be used with a MagPro R30+X100 magnetic stimulator as a medical tool intended to assist in the electrophysiologic monitoring, description and evaluation of muscles, nerves and of structures in the central nervous system in normal and pathological conditions and as an aid for MagVenture TMS Therapy to determine the patient’s individual Motor Threshold (MT). The MEP Monitor is a one channel EMG amplifier mounted on the back of the MagPro, optimized for Motor Evoked Potentials (MEP).
FEATURES
Number of Inputs:
- 1 input protected against electro-static discharge. Balanced inputs.
- 1 pc. 5-pole DIN 240° connector for electrode cable
- Dimensions:
- (HxWxD:) 184 x 94 x 40mm
Weight:
- 0.7kg
Number of Inputs:
- 1 input protected against electro-static discharge. Balanced inputs.
- 1 pc. 5-pole DIN 240° connector for electrode cable
Sound:
- Output for external loudspeaker, 3.5mm jack.
Patient Safety:
- EMG channel galvanically isolated 1.5 kV RMS
Input Impedance:
- 200 MΩ // 100 pF (balanced), >1000 MΩ // 50 pF (common mode)
Noise Level:
- Typical 0.6 µVrms at bandwidth
- 2 Hz to 20 kHz and shorted input
Common Mode Rejection Ratio:
- From surface electrode, through cable and
- amplifier: >55 dB.
- Direct: >100 dB
Isolation Mode Rejection Ratio:
- From input to power ground: >160 dB
Sensitivity Factors:
- 100, 200, 500 µV/Div,
- 1, 2, 5, 10 mV/Div
Time Scales:
- 1, 2, 5, 8, 10 ms/Div
Trigger Mode:
- Level, Autotrig on stim
Lower Frequency Limits (-3dB):
- DSP 1, 2, 5, 10, 20, 50, 100 Hz,
Upper Frequency Limits (-3dB) DSP:
- 1, 2, 5, 10, 20 kHz,
Anti Aliasing:
- 20 kHz (-3 dB), 1st order
Sampling :